This presentation reports a comparison of a handheld Bruker Tracer III XRF and EPA Method 7471A for mercury (Hg) analysis of 51 South River bank soil samples. Calibration used 10 standards (1.07–587 ppm). Method detection limits were ~7.4 ppm (60 s) and ~3.4 ppm (100 s). Median recovery was 96% and mean RSD 23%. Moisture had minimal interference.

File Type: pdf
Categories: Presentations
Tags: bank-soils, bruker-tracer-iii, calibration, field-validation, handheld-xrf, Hg, mercury, method-7471A, moisture-effects, south-river
Publication Date :
Type : Presentation
Review Notes : No publication or presentation date is visible in the PDF. Slide content clearly shows 51 bank soil samples, calibration with 10 standards, use of a Bruker Tracer III handheld XRF, and comparison to EPA Method 7471A (Lancaster Labs referenced).