Slide deck summarizing a 2016 field validation study of handheld XRF for measuring soil mercury. Includes 2015 preliminary results (site-specific calibration R2≈0.996, detection limit ~7.4 ppm, agreement with EPA Method 7471A R2≈0.995) and a 2016 task plan covering field and laboratory validation, accuracy quantification, calibration standards, and optional screening uses.
File Type:
pdf
Categories:
Presentations
Publication Date :
1905-07-08
Type :
Presentation
Review Notes :
Document is a slide presentation authored by Robert Brent, Ph.D., labeled as a 2016 study. Exact publication or presentation date beyond the year is not provided. The acronym 'BMA' appears on slides but is not defined in the PDF. No institutional affiliation for the author is stated on the slides.